What is the purpose of the TruBurst Calibration Plate?
The calibration plate (stock code 550-005) is supplied with TruBurst - James Heal’s Pneumatic Bursting Strength Tester.
The flow control method is used for M&S P27:2010.
- Select the chosen standard and check that the correct test area is being used.
- Set the supply pressure from the compressor to 6 ± 0.2 bar (600 ± 20 kPa or 87± 3 PSI).
- Use the green 777-134 Plain Diaphragms (1mm).
- Place the metal calibration plate between the test dome and the clamp ring / diaphragm.
- Touch “YES” to start automatic flow calibration.
- TruBurst will automatically calibrate to the rate of:
24±1kPa/s M&S P27 Fabric
45 ±1kPa/s M&S P27 Lace