TruBurst

Why do I keep getting Data Logger Run Time Error 75 on TruBurst 600?

When the TruBurst is used in conjunction with a PC data logger the test results on the PC are saved in numerical test result order.

  • If the TruBurst memory is reset (defaulted) then the test count will start off again from 1.
  • If the PC then tries to save test result 1 when it already exists the software generates the run time error 75 message.
  • The solution to this problem is to remove the old rest results from the PC and archive them in a different location.

WARNING: Any maintenance work undertaken on a James Heal instrument must be done so by a trained and competent person. Where electrical work is detailed, this should only be completed by a qualified engineer. James Heal is not responsible for damage or injury arising from work carried out by any person not qualified to do so. If in doubt, please contact James Heal or your local James Heal Agent.